Home > 工業產品 > SITA Process Solutions > sita ConSpector 污染度檢測儀

sita ConSpector 污染度檢測儀

  • Easy control of contamination level in cleaning and rinsing baths
     

     
    Product description
    The SITA ConSpector controls the contamination level in cleaning and rinsing baths. An increasing contamination level causes a decreasing cleaning power. A high parts cleanliness despite of an optimal cleaning agent concentration cannot be guaranteed anymore if the bath is contaminated e.g. due to oils or greases.
     
    The light guide rod at the sensor head of the measuring device is immerged in to the sample to be measured. The measurement starts at the touch of one button. The contamination level is displayed within a few seconds. Based on this measuring result, an optimal point of time for bath care actions or preparation of a fresh cleaning solution can be determined.
     
    Failures of treatment measures can be quickly indentified due to continuous bath replenishment. For an easy operation, the measuring value is compared to an individually determined limit value and when exceeding an alarm signal is sent out by the device.
     
    Controlling the bath contamination ensures a high cleaning quality and thus high parts cleanliness for processes in the field of surface coating and surface treatment. Cleaning bath contamination is reproducibly detected due to the fluorescence measurement. Controlling the contamination level allows reliable and efficient process management.

    Advantages:
    - Determination of contamination level at the touch of one button
    - Targeted initiation and control of bath replenishment actions
    - Identify operation errors e.g. stop of bath treatment actions
    - Automatic evaluation of bath conditions due to pre-determined limit values
    - Use of measuring device independent of used bath chemicals through to a calibration
    - Automatic saving of measuring data including date, time and temperature
     

     
    Measuring principle
    The SITA ConSpector detects the bath contamination such as oils and greases through their fluorescence which is excited by UV light of an LED in the sensor head.
     
    The UV light or fluorescence light is lead into the liquid by a light guide rod at the sensor head. A photodiode in the sensor head measures the fluorescence intensity. A high bath contamination leads to a high measuring value.
     
     

     
    Application
    Removing contamination off cleaning and rinsing bath in the industrial part cleaning industry ensures a long bath operation time and high parts cleanliness. The SITA ConSpector can measure the bath contamination in cleaning and rinsing baths at the plant or in the laboratory. The measuring result is available within a few seconds.
     
    Sudden process disturbances such as the breakdown of the filter unit or a high increase of the contamination level in the bath can be immediately identified and resolved at the same time. Cleaning bath treatment measures can be carried out and controlled according to the application. The cleaning process can be carried out within allowed limit values leading to a reliable cleaning result. Controlling the rinsing steps with regard to the carry-over effect of contamination or cleaning chemicals avoids a further contamination of parts.
     
    Examples for detectable contamination:

    - Cooling lubricants
    - Rolling oil
    - Grease
    - Deep drawing oils
    - Release agents
     

     
    Technical data
    Contamination level / Fluorescence intensity
    Measuring range
    1...40.000 / (1 ... 40.000) RFU*
    Max. measuring deviation
    0.5 % of measuring range
    Liquid temperature
    Measuring range
    (0...100) °C
    Max. measuring deviation
    0,5% des Messbereiches
    Current Supply
    Mains adapter
    5 V/500 mA (USB)
    Battery (Lion, integrated)
    3,6 V/1950 mAh
    Max. power consumption
    ca. 2,5 W
    Excitation diode
    365 nm, max. 150 mW
    Interface, Display, Memory
    USB interface
    Data transfer, power supply & charging of battery
    Display
    LCD, 128 x 64 Pixel,
    illuminated
    Memory
    permanent: measuring parameter / data 255 profiles, 8191 measuring values per profile
    Ambient conditions
    Acceptable ambient temperature (storage / operation)
    (-20...60) °C / (10...50) °C
    acceptable ambient conditions
    rel. humidity< 70 %,
    air pressure(300...1060) hPa
    Housing, Dimensios, Weight
    Housing
    Aluminium, IP 51
    Dimensions (WxHxD)
    Basic Unit
    Sensor head

    129x82x48 in mm
    142x50x30 in mm
    Weight of instrument
    560 g
     * RFU - Relative Fluorescence Unit of SITA-Fluorescence standard
     
 